j*****a
发帖数: 92
1
Amicon ultra-15 centrifugal filter (10 K) 浓缩38 K的蛋白,穿滤. 浓缩液,A3.5 (
280 nm)滤出液 A2.7 (280 nm).是质量不好? 以前也出现过,换一个就好了。这次试了
一盒filter都穿滤.
y******n
发帖数: 8667
2
感觉有的蛋白就浓缩不了。有的会沾膜上?
a*****y
发帖数: 269
3
spead should not > 4000xg
o*****r
发帖数: 156
4
This should not happen.
You know that you should never trust the Abs value when it's more than 2.
1. What is your reference in UV measurement?
2. Do you have anything in the buffer that absorbs at 280nm, for example
imidazole?
3. What't the extension coefficient of your protein?

5 (

【在 j*****a 的大作中提到】
: Amicon ultra-15 centrifugal filter (10 K) 浓缩38 K的蛋白,穿滤. 浓缩液,A3.5 (
: 280 nm)滤出液 A2.7 (280 nm).是质量不好? 以前也出现过,换一个就好了。这次试了
: 一盒filter都穿滤.

m**z
发帖数: 787
5
re this.What is the 280/260 ratio?
i had measured flow through from amicon before and it gave me high 280nm
reading. But when ran on a gel, there is nothing.
I think usually if you don't spin it with higher speed than you supposed to,
it should work just fine.

【在 o*****r 的大作中提到】
: This should not happen.
: You know that you should never trust the Abs value when it's more than 2.
: 1. What is your reference in UV measurement?
: 2. Do you have anything in the buffer that absorbs at 280nm, for example
: imidazole?
: 3. What't the extension coefficient of your protein?
:
: 5 (

o*****r
发帖数: 156
6
Yes, spin at the recommended speed.
We use Amicon very often to concentrate proteins to ~10mg/ml.
I always put blank buffer in it and spin 10-15 mins, then spin my real
sample.
By doing this, the trace amount of glycerol (and other soluble stuff) on the
membrane get washed away.
And you do get a better UV measurement. My flow-thru was never higher than A
0.05 at 280 nm.
Occasionally, the protein might absorb to the membrane, but never ends in
the flow-thru more than 5%.

to,

【在 m**z 的大作中提到】
: re this.What is the 280/260 ratio?
: i had measured flow through from amicon before and it gave me high 280nm
: reading. But when ran on a gel, there is nothing.
: I think usually if you don't spin it with higher speed than you supposed to,
: it should work just fine.

j****s
发帖数: 1366
7
15ml是容易穿,所以我都用4ml的...
s*******e
发帖数: 1010
8
what is in ur protein sample is most important. If it is hi-imidazole
elution, there will lots of small molecules have absorbance at 280nm and
they will go through membrane very easy
1 (共1页)
进入Biology版参与讨论
相关主题
请教IP怎么控制input的量做荧光的tetramer
Ab purification from hybridoma cells纯化蛋白过柱的洗出液内含有很多目标蛋白...
求推荐胶回收试剂盒HIS6-imidazole
IP的elution在跑SDS胶做Mass Spec之前如何有效出去elution sample中的3XFLAG肽
急问, 纯化出来的蛋白怎么保存?请教:BAC Maxi Prep的问题
请蛋白纯化专家指点没有DNA Centrifugal Vacuum怎么浓缩DNA啊?
how to purify the degraded protein bands on the gel?请教几个用来产生 membrane proteins设备的价格 (转载)
Re: His tag的蛋白质不能纯化, 救命[合集] 请求推荐抽BAC clone DNA的kit
相关话题的讨论汇总
话题: amicon话题: 穿滤话题: filter话题: 280nm