j*******2 发帖数: 51 | 1 实验室有个7年旧的shimadzu HPLC,用过约3到4年,然后就闲在那里了。现在需要用它
,发现些问题,想找人做下calibration,maintenace和performance tests,主要目的
是让它重新工作起来,版上有人做相关的服务吗?或有人可以推荐相关的服务机构吗? | S********8 发帖数: 781 | 2 If I were you, I will disassamble all pumps, tubes and parts, then replace
them with new tubes, new parts, leave pumps there if they works well.
Next step after re assambling, manually calibrate the flow rate with water,
MeCN or MeOH, with or without C18 column, and verify the pressure
Check the UV detector to see if the light bubble still works well or replace
it with new light bubble
The reparing costs are very expensive and may allow you to buy a new HPLC
system if you ask experts to do it, So do it by your self
【在 j*******2 的大作中提到】 : 实验室有个7年旧的shimadzu HPLC,用过约3到4年,然后就闲在那里了。现在需要用它 : ,发现些问题,想找人做下calibration,maintenace和performance tests,主要目的 : 是让它重新工作起来,版上有人做相关的服务吗?或有人可以推荐相关的服务机构吗?
| d**********r 发帖数: 24123 | | j*******2 发帖数: 51 | 4 Thanks very much. It really helps.
At the beginning, we were trying to bring a professional here to do this.
But if it's costly, we may consider to do it ourselves. Thanks again.
,
replace
【在 S********8 的大作中提到】 : If I were you, I will disassamble all pumps, tubes and parts, then replace : them with new tubes, new parts, leave pumps there if they works well. : Next step after re assambling, manually calibrate the flow rate with water, : MeCN or MeOH, with or without C18 column, and verify the pressure : Check the UV detector to see if the light bubble still works well or replace : it with new light bubble : The reparing costs are very expensive and may allow you to buy a new HPLC : system if you ask experts to do it, So do it by your self
| S********8 发帖数: 781 | 5 sometime, it is not necessary to keep 4 pumps, most of time (>95%), 2 pumps
(A & B ) are enough, so check 4 pumps and check valve, keep 2 good condition
pumps and tubings in system and remove the other tubings. then make system simple and easy to use |
|